Stratus CS 200 Acute Cardiac Care Troponin Analyzer
Lab-quality results delivered with speed and efficiency where you need them
Stratus CS 200 System* and Assay Specifications
Troponin I Assay | CKMB Assay | NT-proBNP Assay | D-dimer Assay | hsCRP Assay | Myoglobin Assay | βhCG Assay | |
---|---|---|---|---|---|---|---|
Assay Range | 0.03–50 μg/L (30–50,000 ng/L) | 0.3–150 μg/L (ng/mL) | 15–20,000 pg/mL | 6–5000 μg/L (ng/mL) FEU | 0.1–50 mg/L | 1–900 μg/L (ng/mL) | 0.5–1250 IU/L (mIU/mL) |
Analytical Sensitivity | <0.03 μg/L (<30 ng/L) | 0.3 μg/L | 15.0 pg/mL | 6.0 μg/L FEU | ≤0.1 mg/L | 1.0 μg/L | <0.5 IU/L |
Assay Characteristics | 10% at 0.06 μg/L (10% at 60 ng/L) | 4.0% at 3.7 μg/L | 4.4% at 96.6 pg/mL | 4.1% at 412 μg/L | 6.8% at 1.16 mg/L | 3.4% at 56 μg/L; 4.6% at 142 μg/L† | 2.5% at 27.1 IU/L‡ |
Calibration Stability | 60 days | 60 days | 30 days | 60 days | 60 days | 60 days | 90 days |
Please refer to the assay package inserts for more detailed information.
Qualified Collection Tube Types | Becton Dickinson 13 x 75 mm green Hemogard® tube, 4.0 mL draw (lithium or sodium heparin) |
---|---|
Becton Dickinson 13 x 75 mm lt blue Hemogard tube, 4.5 mL draw (3.2% sodium citrate) | |
Sarstedt Monovette® tube 13 x 65 mm orange, 2.6 mL draw (lithium heparin) | |
Sarstedt Monovette tube 13 x 65 mm blue, 2.9 mL draw (sodium citrate) | |
Centrifuge Speed | Microprocessor-verified between 18,000 and 22,000 rpm |
Power Requirements |